7. Description of display symbols
English
Display symbol Status/cause Remedy
n
n
Appears during measure-
ment and flashes once the
pulse is detected.
Ongoing measurement
stay calm!
Display when batteries are
empty.
Replace all 4 batteries
with new ones.
E
ii
-
-1
Pressure raise is defective.
Cuff is not correctly put on
or leaky and/or appliance is
broken.
Check cuff and fixing.
Repeat measurement
after 5 minutes.
E
ii
-
-2
Measurement was incorrect.
User has moved,
is not calm or has the wrong
position.
Stay calm!
Repeat measurement
after 5 minutes.
E
ii
-
-3
Pressure decreases in cuff
too fast.
Cuff is loose or device is
defective.
Put cuff on again.
Repeat measurement
after 5 minutes.
E
ii
-
-4
Abnormal measured values
or appliance is defective.
Maybe, user has
problems with health
and should consult
the doctor.
H I
Overpressure in measure-
ment system.
Tube blocked or device is
defective.
Check tube and cuff.
Repeat measurement
after 5 minutes.
Consult service.
IM_HealthCont.Pro_USB_D-GB-F:IM_Temp Precise_dt-engl-ital 22.12.2009 15:34 Uhr Seite 58