26
CAS ELECTRICS CO.,LTD
First page of Calibration mode
4.1.1 Span Calibration (Menu Code 8110)
(Calibration MENU -> 1. Calibration -> 1. Span Calibration)
*Requires set of certified weights. (For best result prepare 15kg/6kg (max) weights)
Display shows the amount of weight that you will need.
Select “Span Calibration
Empty tray and press “PRINT”
While calibrating zero display shows “Wait4” ~ Wait0” and follow next message for Span
Calibration.
Put on the Weight for Max. Capacities then press “PRINT”
*Menu 8130 sets the max capacity for calibration.
8000
CAL
ModE
CALIBRATION MODE
(1/2)
1. Calibration
2. System Options
3. Printer Hardware
ULoAd
10730
10730
ZERO
CALIBRATION
(
1
/2)
- Remove all weight.
- Press PRINT when ready.
WAit4
10730
10730
ZERO
CALIBRATION
(
1
/2)
- Remove all weight.
- Press PRINT when ready.
LoAd
0
10730
S
PAN CALIBRATION
(2/2)
- Place 15.000 kg on the platter.
- Press PRINT when ready.
WAit4
77407
88137
S
PAN CALIBRATION
(2/2)
- Place 15.000 kg on the platter.
- Press PRINT when ready.