CAS ELECTRICS CO.,LTD
Display shows “Wait4” ~ “Wait0” then following message
Error Message
* When tray is unstable during calibration process, following message appear.
* When calibration weight was too heavy or light, following message will appear.
- Calibration weight limits can re-adjust by menu 8182 “Cal Zero(Span) Max(Min) Range”
* When A/D failure is detected during Calibration following message will appear.
Please check the connector between main board and other controller.
8100
CAL
ModE
CALIBRATION
(1/3)
1. Span Calibration
2. Span/Zero Fine Adjust
3. Capacity & Units
WAit0
2776
2776
Cal Error – Unstable (0x01)
Press Any Key
WAit0
1027
1027
Cal Error – Range Over (0x07)
Press Any Key
WAit0
1027
1027
Cal Error – Wrong ADM (0xff)
Press Any Key