Cleaning, disinfecting and system maintenance are imperative to ensure long life of the suction system. Failure to properly clean, disinfect or maintain the system as per the manufacturer's recommendations may affect the WARRANTY.
This process should be completed each day, after the disinfection process. In most cases, the suction hoses in the dental surgery are part of the dental chair delivery system.
This process should be completed once per week at the end of the working day, after the disinfection process.
Ensure the area around the suction system is clean and clear of debris that may affect ventilation. Check the display for any error messages or warnings.
Performed every 2,000 hours or 12 months. Includes system check, filter o-ring lubrication, and HEPA filter replacement.
Performed every 4,000 hours or 2 years. Includes system check, filter o-ring lubrication, and amalgam retention unit service.
Specific service kits for Micro SMART models, including centrifuge and blower components.
Specific service kits for Turbo SMART models, including centrifuge and blower components.
Alarm messages notify of 95% and 100% amalgam levels. Arrangements for replacement should be made to avoid clinic interruption.
Step-by-step guide for replacing the amalgam container, including sealing the old container and replacing the o-ring.
Step-by-step guide for replacing the amalgam container, including opening drain taps, sealing the old container, and securing the new one.
Step-by-step guide for replacing the amalgam container, including opening drain taps, sealing the old container, and securing the new one.
Cleaning, disinfecting and system maintenance are imperative to ensure long life of the suction system. Failure to properly clean, disinfect or maintain the system as per the manufacturer's recommendations may affect the WARRANTY.
This process should be completed each day, after the disinfection process. In most cases, the suction hoses in the dental surgery are part of the dental chair delivery system.
This process should be completed once per week at the end of the working day, after the disinfection process.
Ensure the area around the suction system is clean and clear of debris that may affect ventilation. Check the display for any error messages or warnings.
Performed every 2,000 hours or 12 months. Includes system check, filter o-ring lubrication, and HEPA filter replacement.
Performed every 4,000 hours or 2 years. Includes system check, filter o-ring lubrication, and amalgam retention unit service.
Specific service kits for Micro SMART models, including centrifuge and blower components.
Specific service kits for Turbo SMART models, including centrifuge and blower components.
Alarm messages notify of 95% and 100% amalgam levels. Arrangements for replacement should be made to avoid clinic interruption.
Step-by-step guide for replacing the amalgam container, including sealing the old container and replacing the o-ring.
Step-by-step guide for replacing the amalgam container, including opening drain taps, sealing the old container, and securing the new one.
Step-by-step guide for replacing the amalgam container, including opening drain taps, sealing the old container, and securing the new one.