CALIBRATION CHAPTER 5
Example: 08.0622 mV
7. Program the measured voltage into the SURETEST Indicator as demonstrated in the example,
“+08.0622” by pushing and holding the Up () or Down () Cursor keys.
8. Press ENTER (). The display momentarily reads:
Saved.
9. Turn the power to OFF. Remove the calibration fixture.
5-13