Do you have a question about the CHF Solutions aquadex FlexFlow and is the answer not in the manual?
Device Type | Ultrafiltration System |
---|---|
Therapy | Ultrafiltration |
Blood Flow Rate | 10 - 40 mL/min |
Display | Touchscreen |
Portability | Portable |
Indication | Fluid overload |
Therapy Mode | Continuous |
Treatment Type | Extracorporeal |
Power Source | AC power |
System indicated for temporary/extended ultrafiltration for fluid overload in patients who failed diuretic therapy and require hospitalization.
Attach cartridge, push until clicks, rotate pump until beeps, then press ACCEPT to enter priming mode.
Disconnect lines from priming bag/UF bag, open clamps, ensure HUB to HUB connection without caps.
Check Pw and Pi pressures, ensure Blood Flow rate is no lower than 20ml/min.
Measures circuit patency on the withdraw (blue) side. Negative pressures -20 to -300 mmHg.
Measures circuit patency on the infusion (clear) side. Positive pressures +20 to +300 mmHg.
Troubleshoot by repositioning, flushing, adjusting flow rate, switching lumens, or changing access.
Measures hemofilter patency. Pressure range is -250 to +200 mmHg.
Measures filter resistance; a value less than 1.5 indicates a viable filter.
Silence alarms, interpret instructions, and press CLEAR to advance or resolve the alert.