4. Servicing and
maintenance
This apparatus does not require any particular maintenance.
Cleaning
Once a year, clean the apparatus using a slightly damp cloth.
Clean the inside of the cells with compressed air.
Temperature check
Each time the temperature setpoint is changed, and at least
every six months, check the measured temperature at the level of
the superior central screw with a precision thermometer ± 1 °C.
If there is a temperature difference during this operation, contact
your CHOPIN Technologies representative.
Fuse replacement
If the apparatus does not switch on and if the power supply has
been checked, replace the EM10 fuse.
Disconnect the apparatus and replace its fuse by using a small
screwdriver.
The fuse is located at the back of the apparatus behind a cover.
Fuse characteristics: 5x20 T 2A 250V.