    Carrie MoodyAug 16, 2025
    What to do if Chroma Power Supply shows Over Temperature Protection (OTP)?
    • D
      Donald AlexanderAug 17, 2025
      If your Chroma Power Supply displays an Over Temperature Protection (OTP) error, it could be due to two reasons: either the ambient temperature is too high, or the vent is blocked. To resolve this, first, ensure that you operate the instrument within a temperature range of 0 to 40°C. Second, check the vent and clear any blockage.
  • T
    Trevor VaughnAug 19, 2025
    What to do if Chroma Power Supply shows Over Current Protection (OCP)?
    • J
      Jennifer HendersonAug 19, 2025
      If your Chroma Power Supply is showing an Over Current Protection (OCP) error, it means the output current exceeds the specified limit or the OCP settings. The solution is to remove the overload or increase the OCP settings.
  • A
    Angela ClarkAug 22, 2025
    What to do if Chroma Power Supply shows Over Power Protection (OPP)?
    • R
      Ronald PetersAug 22, 2025
      If the Chroma Power Supply indicates Over Power Protection (OPP), it means that the output power is exceeding the specified limits. To fix this, remove the overload or increase the OPP settings.
  • N
    Nicole FitzgeraldAug 25, 2025
    What to do if Chroma Power Supply shows Over Voltage Protection (OVP)?
    • B
      Bruce WellsAug 26, 2025
      If your Chroma Power Supply displays an Over Voltage Protection (OVP) error, it indicates that the output voltage exceeds the specified limits or the OVP settings. First, check the OVP settings. If the issue persists, consult with your local sales agent.
  • N
    Nicole BrownAug 28, 2025
    Why is my Chroma Power Supply showing Input Error Protection 1 AC FAULT?
    • K
      Kevin MaciasAug 28, 2025
      The Chroma Power Supply shows 'Input Error Protection 1 AC FAULT' because the voltage of the AC input line is either too low or too high. To resolve this, measure the input voltage and adjust it if it exceeds the specified limits.
  • E
    Elizabeth HarrisAug 30, 2025
    How to fix D/D power stage error protection (D/D FAULT) on Chroma Power Supply?
    • M
      maldonadoamyAug 30, 2025
      If the D/D FAULT protection is triggered on your Chroma Power Supply, it could be due to a transient current that is too big or the D/D power stage being damaged. When this occurs, first turn off the power supply and remove the load. Make sure the cables are connected correctly, then power it on again.
  • S
    Shannon ChaneySep 2, 2025
    What to do if Chroma Power Supply has no output voltage?
    • N
      Noah AllenSep 2, 2025
      If your Chroma Power Supply has no output voltage, it could be due to the output voltage feedback being abnormal or the D/D power stage being damaged. Consult with your local sales agent if it is unable to reset the protection state.
  • S
    Sean RiveraSep 5, 2025
    How to fix Fan Fail Protection (FAN LOCK) on Chroma Power Supply?
    • C
      Cheryl LeeSep 5, 2025
      If your Chroma Power Supply displays 'Fan Fail Protection (FAN LOCK)', it means that the fan is out of order or the feedback circuit is abnormal. Consult with your local sales agent if it is unable to reset the protection state.
  • T
    Thomas HaydenSep 7, 2025
    Why is Chroma Power Supply output not within Accuracy SPEC?
    • A
      alec36Sep 7, 2025
      The Chroma Power Supply output may not be within the accuracy specification due to feature swings caused by aged components. It requires periodic calibration. See section 3.2.4.4 Calibration.
  • C
    coxveronicaSep 10, 2025
    Why does Chroma Power Supply have bad measurement for V, I?
    • L
      Lisa AlexanderSep 12, 2025
      The Chroma Power Supply might show bad measurements for Voltage (V) and Current (I) because of feature swings due to aged components. It needs calibration periodically. See section 3.2.4.4 Calibration.

Summary

Warranty Information

Safety Summary

General Safety Precautions

Essential safety precautions for operation, service, and repair of the instrument.

Overview of 62000 D Series

System Functions Overview

Details the main functions of the DC power supply.

Specifications Overview

Lists the technical specifications of the 62000D Series models.

Installation Guide

Preparation for Use

Remote Sensing Connection

Power On Procedure

Manual Operation Guide

Menu Navigation and Setup

Explains how to use the setup menu.

Meas. & Setup Menu

System Setup Menu

Protection Functions Overview

OVP (Over Voltage Protection) Setup

OCP (Over Current Protection) Setup

OPP (Over Power Protection) Setup

Safety Interlock Function

Foldback Protection Settings

Configuration Menu

Series and Parallel Connection

Setting Series;Parallel Operation Mode

Series Mode Connection Setup

Calibration Procedures

Voltage Calibration

Steps for calibrating voltage output and measurement.

Current Calibration Procedures

External Output Settings

Program Sequence Operations

List Mode Operation

Programming sequences in List Mode.

V Step Mode Operation

Programming sequences in V Step Mode.

List Mode Programming

PROGRAM Settings

Configures program settings like trigger, count, and chain.

Setting Program Sequence Details

V Step Mode Operation

Setting V Step Mode Parameters

Configures parameters for V Step Mode.

V Step Mode Execution

Remote Operation Overview

USB Interface Details

Information on USB connectivity and driver support.

Setting GPIB and Ethernet Parameters

Configuring GPIB and Ethernet for remote control.

DC Power Supply Commands

Common Command Syntax

Explains command structure, conventions, and data formats.

Specific Commands for 62000 D Series

CONFIGURE Commands

SOURCE Subsystem Commands

LOAD Subsystem Commands

FETCH Subsystem Commands

MEASURE Subsystem Commands

PROGRAM Subsystem Commands

SYSTEM Subsystem Commands

INSTRUMENT Subsystem Commands

Self Test and Troubleshooting

Troubleshooting Common Problems

Solutions for common operational issues and error messages.

List of System Protections

Chroma 62120D-1200 Specifications

General IconGeneral
BrandChroma
Model62120D-1200
CategoryPower Supply
LanguageEnglish

Related product manuals