    drobertsAug 19, 2025
    What to do if Chroma 62150H-600S shows Over Temperature Protection (OTP)?
    • S
      Sandra GutierrezAug 19, 2025
      If the Chroma Power Supply displays an Over Temperature Protection (OTP) error, it could be due to two reasons. First, the ambient temperature might be too high. Ensure the instrument is operated within a temperature range of 0 to 40°C. Second, the vent could be blocked. Make sure to clear the vent to allow for proper ventilation.
  • M
    Margaret ConleyAug 21, 2025
    What to do if Chroma Power Supply shows Over Current Protection (OCP)?
    • V
      vgomezAug 21, 2025
      If the Chroma Power Supply shows an Over Current Protection (OCP) fault, it means the output current exceeds the specified limits or the OCP settings. To resolve this, remove the overload connected to the power supply or increase the OCP settings.
  • M
    Marcus StewartAug 23, 2025
    What to do if Chroma 62150H-600S shows Over Power Protection (OPP)?
    • J
      Johnny JonesAug 23, 2025
      If the Chroma Power Supply displays an Over Power Protection (OPP) error, it indicates that the output power is exceeding the specified limits or the OPP settings. To fix this, remove the overload or increase the OPP settings.
  • J
    Jonathan PayneAug 26, 2025
    What does Input Error Protection 1 AC FAULT mean for Chroma 62150H-600S?
    • S
      Sharon PetersonAug 26, 2025
      If you encounter an Input Error Protection 1 AC FAULT on your Chroma Power Supply, it means that the voltage of the AC input line is either too low or too high. To solve this, measure the input voltage and adjust it if it exceeds the specified limits.
  • T
    Taylor NormanAug 28, 2025
    What to do if Chroma 62150H-600S shows Over Voltage Protection (OVP)?
    • J
      John WatsonAug 28, 2025
      If the Chroma Power Supply shows an Over Voltage Protection (OVP) fault, it means the output voltage exceeds the specified limits or the OVP settings. Check the OVP settings. If the protection state cannot be reset, contact your local sales agent for further assistance.
  • A
    alicia18Aug 30, 2025
    What does D/D FAULT mean on Chroma 62150H-600S Power Supply?
    • J
      Jose BarberAug 30, 2025
      If the Chroma Power Supply displays a D/D FAULT error, it could be due to a large transient current or damage to the D/D power stage. If a D/D FAULT protection occurs, first turn off the power supply and remove the load. Ensure that the cables are connected correctly, and then power it on again. If the issue persists, contact your local sales agent for further assistance.
  • J
    John CollinsSep 1, 2025
    Why does my Chroma 62150H-600S show 'ERROR!!! MASTER OR SLAVE NO MATCH'?
    • G
      grace33Sep 1, 2025
      If your Chroma Power Supply shows 'MATCH warning that the models are incompatible when connected in series or parallel (ERROR!!! MASTER OR SLAVE NO MATCH)', this means the model numbers do not match. Different model power supplies cannot be connected in series or parallel mode. If you need further assistance, contact your local sales agent.
  • H
    Heather BellSep 4, 2025
    Why does my Chroma 62150H-600S Power Supply have no output voltage?
    • C
      Cory NelsonSep 4, 2025
      If the Chroma Power Supply is not providing any output voltage, it could be due to abnormal output voltage feedback or damage to the D/D power stage. If the protection state cannot be reset, contact your local sales agent.
  • C
    Cindy PenningtonSep 6, 2025
    How to fix Chroma DC Power Supply control issues via GPIB?
    • L
      Lauren WalkerSep 6, 2025
      If you are unable to control the Chroma DC Power Supply via GPIB, it could be due to an incorrect address or a loose/disconnected GPIB cable. Update the address and check the cable connection, securing it with screws.
  • J
    julie60Sep 8, 2025
    What does 'Current sharing cable connection error (C/S CABLE ERR.)' mean on Chroma 62150H-600S Power Supply?
    • J
      john01Sep 8, 2025
      If the Chroma Power Supply displays 'Current sharing cable connection error (C/S CABLE ERR.)', it means the current sharing cable is connected when used in series mode. Check if the cables are connected correctly. Contact your local sales agent for further assistance.

Chroma 62150H-600S Specifications

General IconGeneral
Model62150H-600S
CategoryPower Supply
Output Voltage0-600 V
Number of Outputs1
DisplayLCD
Operating Temperature0°C to 50°C
Storage Temperature-20°C to 70°C
Output Current0-25 A
InterfaceRS-232, USB, GPIB, Ethernet
Humidity≤80% RH, non-condensing
Voltage Resolution1 mV
Current Resolution1mA