Remote Operation
5-33
5.6.2.3 SOURCE Subsystem
1. SOURce:VOLTage
Description: Sets the output voltage.
Syntax: SOURce:VOLTage <NRf+>[suffix]
SOURce:VOLTage <NRf+>[suffix]
Parameter: Refer to individual spec for valid numeric range.
Example: SOUR:VOLT 0.01 Sets the output voltage to 0.01 volt.
SOUR:VOLT 80.00 Sets the output voltage to 80.00 volt.
Query Syntax: SOUR:VOLT?
Return Parameter: <NRf+> [Unit Volt]
Query Example: SOUR:VOLT? Returns the output voltage setting value.
Return Example: 8.000000e+01
2. SOURce:VOLTage:LIMit:{HIGH/LOW}
Description: Sets the output voltage range.
Syntax: SOURce:VOLTage:LIMIT:HIGH <NRf+>[suffix]
SOURce:VOLTage:LIMIT:LOW <NRf+>[suffix]
Parameter: Refer to individual spec for valid numeric range.
Example: SOUR:VOLT:LIMIT:HIGH 60.0 Sets the output voltage range to
60V maximum.
SOUR:VOLT:LIMIT:LOW 20.0 Sets the output voltage range to
20V minimum.
Query Syntax: SOUR:VOLT:LIMIT:HIGH?
SOUR:VOLT:LIMIT:LOW?
Return Parameter: <NRf+> [Unit Volt]
Query Example: SOUR:VOLT:LIMIT:HIGH? Returns the maximum range
setting for voltage.
Return Example: 8.000000e+01
3. SOURce:VOLTage:PROTect:{HIGH}
Description: Sets the voltage range for over voltage protection.
Syntax: SOURce:VOLTage:PROTect:HIGH <NRf+>[suffix]
Parameter: Refer to individual spec for valid numeric range.
Example: SOUR:VOLT:PROT:HIGH 60.0 Sets the high limit to 60V for
voltage output protection.
Query Syntax: SOUR:VOLT:PROT:HIGH?
Return Parameter: <NRf+> [Unit Volt]
Query Example: SOUR:VOLT:PROT:HIGH? Returns the high limit setting for
voltage protection.
Return Example: 8.800000e+01
4. SOURce:VOLTage:SLEW
Description: Sets the rising or falling slew rate (volt/ms) of the output voltage.
Syntax: SOURce:VOLTage:SLEW <NR1>[suffix]
SOURce:VOLTage:SLEW <NR1>[suffix]
Parameter: Refer to individual spec for valid numeric range.
Example: SOUR:VOLT:SLEW 0.01 Sets the output voltage slew rate to
0.01volt/mS
SOUR:VOLT:SLEW 10 Sets the output voltage slew rate to
10 volt/mS
Query Syntax: SOUR:VOLT:SLEW?
Return Parameter: <NR1> [Unit Volt/ms]