Remote Operation
5-27
Table 5-8
Bit No.
Bit Weight
Description
7 128
Power on Bit. Rebooting the Power Supply will set
this bit to 1.
6
Always 0.
5 32
Command Error Bit. This bit is set to 1 if there is
any IEEE 488.2 syntax error.
4 16
Execution Error Bit. This bit is set to 1 when the
command parameter is out of valid range or
inconsistent.
3 8
Device Dependent Error Bit. This bit is set to 1
when too many errors have occurred and
the error
queue is full.
2 4
Queue Error Bit. This bit is set to 1 when reading
data from the output buffer and no data is present,
or when the data is lost.
1
Always 0.
0
1
5.6.2 Specific Commands for 62000H Series
5.6.2.1 ABORT Subsystem
ABORt
Description: Sets all output states to “OFF”.
Syntax: ABORt
5.6.2.2 CONFIGURE Subsystem
1. CONFigure:BEEPer
Description: Sets the beeper to ON or OFF.
Syntax: CONFigure:BEEPer ON
CONFigure:BEEPer OFF
Parameter: ON|OFF
Example: CONF: BEEPer ON
CONF: BEEPer OFF
Query Syntax: CONFigure:BEEPer?
Return Parameter: ON | OFF
Query Example: CONF:BEEPer? Returns the beeper control status.
Return Example: ON or OFF
2. CONFigure:REMote
Description: Sets the remote control status (valid for RS232C only).
Syntax: CONFigure:REMote ON
CONFigure:REMote OFF
Parameter: ON|OFF
Example: CONF:REM OFF Disables remote control.