  • R
    robert08Aug 18, 2025
    What to do if Chroma 62020H-150S Power Supply shows Over Temperature Protection (OTP)?
    • M
      Monica LeeAug 18, 2025
      If the Chroma Power Supply displays an Over Temperature Protection (OTP) error, it could be due to two reasons. First, the ambient temperature might be too high. Ensure the instrument is operated within a temperature range of 0 to 40°C. Second, the vent could be blocked. Make sure to clear the vent to allow for proper ventilation.
  • A
    Andre PetersenAug 20, 2025
    What to do if Chroma 62020H-150S shows Over Current Protection (OCP)?
    • P
      Paul DixonAug 20, 2025
      If the Chroma Power Supply shows an Over Current Protection (OCP) fault, it means the output current exceeds the specified limits or the OCP settings. To resolve this, remove the overload connected to the power supply or increase the OCP settings.
  • L
    Lindsey Clark DVMAug 23, 2025
    What to do if Chroma 62020H-150S shows Over Power Protection (OPP)?
    • R
      Ruth WernerAug 23, 2025
      If the Chroma Power Supply displays an Over Power Protection (OPP) error, it indicates that the output power is exceeding the specified limits or the OPP settings. To fix this, remove the overload or increase the OPP settings.
  • S
    sean41Aug 26, 2025
    What does Input Error Protection 1 AC FAULT mean for Chroma 62020H-150S Power Supply?
    • K
      Katherine RussellAug 26, 2025
      If you encounter an Input Error Protection 1 AC FAULT on your Chroma Power Supply, it means that the voltage of the AC input line is either too low or too high. To solve this, measure the input voltage and adjust it if it exceeds the specified limits.
  • D
    Dr. Kelly ProctorAug 27, 2025
    What to do if Chroma 62020H-150S shows Over Voltage Protection (OVP)?
    • L
      Laura WilliamsAug 28, 2025
      If the Chroma Power Supply shows an Over Voltage Protection (OVP) fault, it means the output voltage exceeds the specified limits or the OVP settings. Check the OVP settings. If the protection state cannot be reset, contact your local sales agent for further assistance.
  • T
    Thomas RuizAug 30, 2025
    What does D/D FAULT mean on Chroma 62020H-150S Power Supply?
    • C
      Caleb SmithAug 30, 2025
      If the Chroma Power Supply displays a D/D FAULT error, it could be due to a large transient current or damage to the D/D power stage. If a D/D FAULT protection occurs, first turn off the power supply and remove the load. Ensure that the cables are connected correctly, and then power it on again. If the issue persists, contact your local sales agent for further assistance.
  • M
    Mrs. Katherine Peters MDSep 1, 2025
    Why does my Chroma 62020H-150S Power Supply show 'ERROR!!! MASTER OR SLAVE NO MATCH'?
    • R
      Ronald PatelSep 1, 2025
      If your Chroma Power Supply shows 'MATCH warning that the models are incompatible when connected in series or parallel (ERROR!!! MASTER OR SLAVE NO MATCH)', this means the model numbers do not match. Different model power supplies cannot be connected in series or parallel mode. If you need further assistance, contact your local sales agent.
  • S
    Sheila AustinSep 3, 2025
    Why does my Chroma 62020H-150S Power Supply have no output voltage?
    • H
      Heidi ReynoldsSep 3, 2025
      If the Chroma Power Supply is not providing any output voltage, it could be due to abnormal output voltage feedback or damage to the D/D power stage. If the protection state cannot be reset, contact your local sales agent.
  • B
    Barbara DavisSep 6, 2025
    How to fix Chroma DC Power Supply control issues via GPIB?
    • L
      Latoya EspinozaSep 6, 2025
      If you are unable to control the Chroma DC Power Supply via GPIB, it could be due to an incorrect address or a loose/disconnected GPIB cable. Update the address and check the cable connection, securing it with screws.
  • D
    Daniel PowellSep 8, 2025
    What does 'Current sharing cable connection error (C/S CABLE ERR.)' mean on Chroma Power Supply?
    • M
      Michael FreemanSep 8, 2025
      If the Chroma Power Supply displays 'Current sharing cable connection error (C/S CABLE ERR.)', it means the current sharing cable is connected when used in series mode. Check if the cables are connected correctly. Contact your local sales agent for further assistance.

Chroma 62020H-150S Specifications

General IconGeneral
Model62020H-150S
CategoryPower Supply
Output Power3000 W
DisplayLCD
CoolingForced Air Cooling
Voltage Resolution1 mV
Current Resolution1 mA
Operating Temperature0°C to 40°C
Voltage Accuracy0.05% + 10mV
Current Accuracy0.1% + 10mA
InterfaceRS232, USB, GPIB
Storage Temperature-20°C to 70°C
Relative Humidity20% to 80%