CALIBRATION:
Your balance should be calibrated on a regular basis for the best performance. The most accurate
calibration is made only after the balance has been connected to a power source for 30 minutes.
Under continuous use, calibrate the balance at least every 30 days, any time it is moved or bumped,
and whenever the ambient temperature changes by more than 3 degree Celcius.
MODEL
CALIBRATION WEIGHT
MJ-300
3009 or 2009
MJ-500
5009 or 3009
MJ-3000
30009 or 20009
MJ-6000
60009 or 50009
CALIBRA TlON PROCEDURE:
···Check the spirit level of the balance and adjust if
necessary .
.. ·Make certain the weighing pan is empty .
I CAL
I CAL
o
.. ·Press the F key until "CAL" appears on the display.
.. ·Press the TARE key to start calibration .
.. ·Press the TARE key.
... The balance automatically sets ZERO.
(In case of MJ-3000 1
CAL
OR
CAL
I PASS
3 O O O
2 O O O
... The balance will then request the appropriate size
calibration weight.
Carefully place the calibration weight on the weighing
pan.
.. ·Press the U key.
.. ·Press the TARE key.
... " PASS" appears on the display when calibration is
complete.
-ll-