Troubleshooting
NIBP fault
Possible Reason
It can be inflated, but the NIBP value
cannot be measured. Sometimes
inflation is repeating, prompts: time
out
Inappropriate NIBP settings
-inflation value, patient type
Incorrect NIBP operation
Choose the right cuff, and ensure patient stability
The cuffs are old or of poor quality
Check the quality of the cuff and replace it with a new
one
NIBP cuff, tubing, pump, or valves have air leak
Do the NIBP leak test, check the leak location, and
replace the leaking parts
Parameter module failure
Replace the parameter module
The measured NIBP value is inaccurate
and the deviation is large
NIBP operation, inappropriate patient status
Check measurement operation, patient status, and
ensure correct operation and stable status
NIBP measurement sensor miscalibration
Perform NIBP calibration and verification
Start NIBP, inflation is not started
Air pump failure
Check the connection of the NIBP air pump, replace the
air pump
Parameter module failure
Replace the parameter module