Troubleshooting
IBP fault
Fault Phenomenon
Possible Reason
Inspection and Repair
IBP waveform and value are not
accurate
IBP zeroing not performed
Zero calibration is required after powering on or
replacing the IBP sensor
IBP sensor or IBP cable fault
Replace accessories
Parameter module failure
Replace the parameter module
IBP waveform and parameter area are
not displayed, sometimes there are
technical alarms
The IBP function is not turned on, and the IBP waveform
display is not turned on
Turn on IBP in the module switch settings and turn on
the IBP waveform area in the screen display settings
IBP sensor or IBP cable fault
Replace accessories