Troubleshooting
EtCO2 fault
Inaccurate measurement of EtCO2
No CO2 zeroing before use
Perform zero calibration of the EtCO2 probe
Incorrect settings such as altitude
Check EtCO2 settings to ensure proper settings
Replace accessories such as sampling tube, drying tube,
airway adapter, etc.
The CO2 waveform and parameter
area are not displayed, sometimes
there are technical alarms
The CO2 function is not turned on, and the CO2
waveform display is not turned on
Turn on CO2 in the module switch settings, and turn on
the CO2 waveform area in the screen display settings
Inaccurate CO2Type setting in factory maintenance
Check the CO2Type setting to make sure it matches the
CO2 module in using