5.1 Main unit
Cleaning may be required as a result of spilled drops of blood or when required after
inspection.
If spillage occurs, the unit must immediately be removed from service and cleaned com-
pletely before resuming use.
Use a soft lint-free tissue, moistened with a mild detergent to clean the outside of the
Main unit.
5.2 Hand Unit
Cleaning may be required as a result of spilled drops of blood or when required after
inspection.
24
CS524 Qseal-auto Operator´s manual
❷
❶
For cleaning the electrodes you should loosen the front side of
the handle, by gently turning the holder ring anti-clockwise ❶.
Pull out and remove the fast electrode. Clean the tube and both
electrodes with a soft lintfree tissue moistened with mild de-
tergent. Dry carefully and ensure that the electrodes are com-
pletely dry to prevent sparks. To clean the piston shaft, use a
dry cotton swab.
Note! After cleaning inspect the electrode for damage.
Assemble the parts in reverse order. Mount the electrode en-
suring that it is positioned so as to be parallel with the moving
electrode ❷. Turn the holder-ring fully clockwise and ensure
that the cover is well tightened.
Note! We recommend some sealing tests before resuming
use, ensure that the electrode moves smoothly in
the hand unit.
Caution! Do not dip the units in liquid, as they are not water-
proof. Intruding liquid will cause malfunction, tiny
arcs and reduce the lifetime use.