CS524 Qseal-auto Operator´s manual
Problem Probable cause Recommended action
The sealing
doesn’t start.
Wet or dirty electrodes. Dry/clean the electrodes, see chapter 5.2
No ”click” when you press
the SEAL-button.
Change Hand unit.
No connection. Check coaxial cable connections.
Coaxial cable broken. Change coaxial cable.
Other cause. Contact your local C.M. representative.
Incomplete sealing Low air pressure. Check Coaxial cable connection.
Low voltage. Check battery charge level.
Other cause. Contact your local C.M. representative.
Intermittent sealing Coaxial cable broken. Change coaxial cable or contact your local
C.M. representative.
Seal lamp lights
after seal switch
has been released
Seal switch is broken. Contact your local C.M. representative.
Seal lamp
doesn’t light
Hand unit inoperative. Contact your local C.M. representative.
Coaxial cable broken. Contact your local C.M. representative.
Seal lamp ickers Coaxial cable broken. Change coaxial cable or contact your local
C.M. representative.
Seal lamp doesn’t
light but electrode
moves
Main unit inoperative. Contact your local C.M. representative.