Normal Operation "Feed Only Set" Directions For Use
Battery Operation
Unplugging the pump from the AC line will automatically place the pump on battery power. To
recharge the battery, plug the pump into an AC wall outlet. Pump will automatically begin charging.
General Instuctions
1. Fill or spike feeding container prior to setting up pump.
2. Turn pump on - Press
Power button in lower right-hand corner.
3. In order to achieve proper accuracy, the fluid line in the feeding set bag must be 6
inches above the top of the feeding pump, when initiating the feeding cycle.
4. Select
"Keep Settings" or "Clear Settings" (for day to day use with the same feeding
regimen the
"Keep Settings” option should be selected).
5. Load the Feeding Set per diagram.
Do not overstretch the tubing.
Priming the Pump
1. Press "Prime Pump" to access the pump priming options.
2. Press
"Auto Prime" to automatically prime the pump set. The pump will quickly prime the
feed line and stop the formula before reaching the end of the feeding line.
3. Press the
"Hold To Prime Feed" menu selection to manually top off the line.
4. Press
"Done".
Setting the Feed Rate
1. Select "Adjust Feed" then "Feed Rate" use the buttons on the left to program the pump
from 1 to 400 in increments of 1ml. Select
"Enter" when desired rate is set.
Note the Volume To Be Delivered "VTBD" Rate - is an Optional Feature
Only use the “VTBD” option if you want the pump to stop and alarm once a set amount of
formula is delivered.
Select "Run" you will notice a small drop scroll down the screen when running.
Re-Priming the Pump After the Feed Bag Empties
1. A pump set bag that has been emptied will trigger the Feed Error screen. In this condition the
pump set bag can be refilled to continue the feeding, but only after the pump set has been re-primed.
2. Disconnect the feeding line from the patient.
3. Refill the bag.
4. Press
"Continue" to begin the pump running.
5. Press
"Hold", then press "Adjust Settings", then press "Prime Pump".
6. Press
"Hold To Prime" to prime the line. (Do Not Use Auto Prime)
7. Press
"Done", then select "Run"
To Change Rate or Clear Volume
1. Select "Hold".
2. Select
"Clear Vol Fed" to clear the volume.
3. Select
"Adjust Settings" to adjust all settings.
4. Select
"Run" to return to normal operations.
Pump sets should not be reused after 24 hours of initial usage.
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
t
Enteral Feeding Pump
Kangaroo
Joey

Other manuals for Coviden Kangaroo Joey

Questions and Answers

  • A
    Amanda AlexanderAug 14, 2025
    What to do if Coviden Medical Equipment shows a 'Battery Low' error?
    • K
      Kathleen DavisAug 14, 2025
      If the Coviden Medical Equipment displays a 'Battery Low' error, it means the battery charge is too low. Immediately plug the pump into an A/C outlet to recharge the battery and provide power.
  • A
    Albert LeeAug 17, 2025
    What to do if Coviden Kangaroo Joey Medical Equipment displays a 'Flow Error'?
    • J
      Jenna DeckerAug 17, 2025
      If the Coviden Medical Equipment shows a 'Flow Error', it indicates an occlusion in the pump-patient line (downstream occlusion). Check the line between the pump and the patient for any obstructions and correct them. If the error persists, replace the Pump Set.
  • L
    Lisa GarciaAug 21, 2025
    What to do if Coviden Medical Equipment displays a 'Feed Error'?
    • S
      Sharon MurilloAug 21, 2025
      If the Coviden Medical Equipment displays a 'Feed Error', it indicates an empty bag or an occlusion in the bag-pump line (upstream occlusion). Check if the bag is empty and refill it, or check for and correct any occlusion in the line between the pump and the feed bag. If the error persists, replace the Pump Set.
  • T
    Tammy SanchezAug 25, 2025
    What to do if Coviden Medical Equipment says 'Pump Set Dislodged'?
    • B
      Benjamin ThompsonAug 26, 2025
      If the Coviden Medical Equipment displays 'Pump Set Dislodged', it means the Pump Set is not properly loaded. Check the black ring retainer (MISTIC) and valve; then, reload the Pump Set and restart the pump. If the error persists, replace the Pump Set.
  • K
    Kimberly RuizAug 28, 2025
    What does 'Rotor Error' mean on Coviden Kangaroo Joey Medical Equipment?
    • V
      vmendezAug 28, 2025
      If the Coviden Medical Equipment shows a 'Rotor Error', it indicates that the Pump Set tubing is not properly loaded on the rotor, or there is some other unusual rotor operating condition. Check that the tubing is correctly loaded on the rotor and that it is not damaged or torn. Reload the tubing or replace the Pump Set.
  • K
    Kelly EllisSep 1, 2025
    What does 'Flow Error Use > 24 Hrs' mean on Coviden Kangaroo Joey Medical Equipment?
    • T
      Tyler SmithSep 1, 2025
      If the Coviden Medical Equipment displays a 'Flow Error Use > 24 Hrs' message, it means the Pump Set has been in use beyond the recommended time. This informational warning message will blink on the upper left of the screen, indicating that it is recommended to replace the Pump Set.
  • M
    Molly WrightSep 5, 2025
    How to fix a 'Hold Error' on Coviden Kangaroo Joey Medical Equipment?
    • T
      Tyler SnyderSep 5, 2025
      The 'Hold Error' on Coviden Medical Equipment means that the pump has been in HOLDING mode for more than 5 minutes. Press CONTINUE to return to the HOLDING screen, or press and hold the Power button to turn off the device. From the HOLDING screen, you have options to RUN or ADJUST SETTINGS.
  • D
    Donald SerranoSep 8, 2025
    What to do for 'System Error' on Coviden Kangaroo Joey Medical Equipment?
    • E
      emendozaSep 8, 2025
      To address a 'System Error' on the Coviden Medical Equipment, power down the pump and attempt to restart it to clear the error.

Summary

Normal Operation Feed Only Set Directions For Use

Battery Operation

Explanation of pump operation on battery power and charging procedures.

General Instructions

Basic setup and operational steps for the enteral feeding pump.

Priming the Pump

Step-by-step guide for priming the feeding set and pump.

Setting the Feed Rate

Instructions for programming the pump's feeding rate.

Re-Priming the Pump After the Feed Bag Empties

Procedure for re-priming the pump when the feed bag is empty.

To Change Rate or Clear Volume

Guidance on adjusting the feeding rate or clearing the fed volume.

Troubleshooting Guide

Hold Error

Troubleshooting steps for pump being in HOLDING mode for over 5 minutes.

Flow Error

Diagnosing and correcting occlusions in the pump-patient line.

Feed Error

Resolving issues related to an empty or occluded feed bag line.

Pump Set Dislodged

Steps to address a pump set that is not properly loaded.

Battery Low

Actions to take when the pump's battery charge is too low.

Rotor Error

Troubleshooting pump set tubing loading on the rotor.

System Error

General error resolution, including contacting customer service.

Coviden Kangaroo Joey Specifications

General IconGeneral
TypeEnteral Feeding Pump
ModelKangaroo Joey
Battery TypeRechargeable Lithium-ion
Alarm SystemAudible and visual alarms
Power Requirements100-240 VAC, 50/60 Hz
UsageEnteral feeding
Flow Rate Range1 to 400 mL/hr
Keep Vein Open (KVO) Rate1 mL/hr
Power SourceAC power or battery
Operating Temperature10°C to 40°C (50°F to 104°F)
DisplayLCD