WARNINGS!
Before sterilizing any articles, make sure you first check with their manufacturer if they are autoclavable (resistant to
the temperature of 135ºC and the presence of steam and pressure);
Do not allow patients or especially children to get close to the autoclave;
Install the autoclave in an exclusive sterilization room;
Never warm up or sterilize any kind of food in the autoclave;
Never make any kind of experiment with animals in the autoclave;
Never make any kind of use of this equipment other than the ones described in this manual.
SPECIAL MEASURES AND OBSERVATIONS DURING THE USE OF THE AUTOCLAVE:
Always make sure the door handle is properly locked before switching the autoclave on. See “How to Use the
Autoclave Vitale”, (Item c, page 13). Not following this procedure may cause the door gasket to pop off;
When unlocking the handle, the autoclave door should open easily. Always check for full depressurization, only the
first LED (position zero) of the indicative LEDs bar (Fig. 5, page 10) must be on. Never force the door to open the
autoclave!;
It is normal that some steam comes out through the door when it is opened at the end of the depressurization for
the drying stage;
Never touch the external steam exit (Fig. 1, page 8) and/or the internal surfaces of the autoclave (chamber, trays,
materials, etc.) during or right after operation, they will be very hot. Even after waiting for the materials to cool down,
it is recommended to use proper gloves to handle the sterilized materials. Improper use may cause burns,
Cristófoli is not responsible for incorrect procedures that may cause accidents;
In case of activation of one of the safety devices (sudden steam escape), generally caused by obstruction of the
steam exit internal orifice or by an obstruction of the solenoid valve, wait for the complete depressurization before
opening the door;
IMPORTANT! Always make sure you unplug the autoclave before performing any kind of maintenance like changing
the fuse or even the everyday cleaning;
We recommend reading this manual until it is fully understood. Keep it at hand and use it as a constant reference
source.
IMPORTANT SAFETY INFORMATION
Before using your autoclave Vitale, It is necessary to observe some safety measures. Autoclaves are equipment
which work at high temperature and pressure, therefore they must be handled by qualified and well-informed
personnel, regarding their features and functioning. It is essential for such qualification that the operators read all the
instructions carefully before using the autoclave to make sure to understand them correctly. The intended use of this
equipment is to perform sterilization on dental, medical and hospital instruments/articles resistant to the temperature of
135ºC, steam and pressure.
ATTENTION! TO OPEN THE AUTOCLAVE
1 - Press the door forward; 2 - Lift the handle; 3 - Slide it to the left.
CRISTÓFOLI
A U T O C L A V E S
I N S T R U C T I O N M A N U A L