CRISTÓFOLI’S MISSION
CRISTÓFOLI QUALITY AND ENVIRONMENTAL POLICY
Autoclave Vitale 12/21
Cristófoli, Brazilian company importer and manufacturer of health products
certified by ISO 9001 - Quality Management System, ISO 13485 - Medical
Devices - Quality Management System - Requirements for Regulatory Purposes,
ISO 14001 - Environmental Management and *BPF - Boas Práticas
de Fabricação - ANVISA.
13485
D
E
S
G
A
E
M
S
E
T
T
Ã
S
O
I
S
*BPF - Boas Práticas de Fabricão: Brazilian standard similar to the GMP - Good Manufacturing Practices (FDA / US).
Develop innovative solutions to protect life and promote health.
Cristófoli Equipamentos de Biossegurança Ltda., established at Rodovia BR-158, 127, Jardim Curitiba in Campo
Mourão, Paraná, Brasil, manufactures biosafety equipment to assist the health field having as policy: “Develop innovative
solutions for the health field by using agile, robust and objective processes to better serve its clients. Fulfill the requirements
for regulatory purposes of the applicable standards, promote the continuous improvement of its quality and environmental
systems, prevent pollution, reduce its environmental impacts and continuous training of its employees, achieving this way, a
sustainable profitability and the maximization of the company’s value”. Rev. 2.

Other manuals for Cristofoli vitale 21

Questions and Answers

  • C
    Christopher StephensonJul 28, 2025
    How to fix Cristofoli vitale 21 Medical Equipment if the autoclave does not switch on?
    • V
      Virginia SmithJul 28, 2025
      If your Cristofoli Medical Equipment autoclave doesn't switch on, first check for a power outage in your area. Then, ensure the power cord is properly connected to the outlet. Press the power button to turn it on. If it still doesn't work, replace the fuse located near the power cord: unplug the autoclave, remove the fuse-holder cover, remove the damaged fuse, insert a new one, replace the cover, and turn on the equipment.
  • A
    Ashley TaylorAug 4, 2025
    What to do if Cristofoli vitale 21 autoclave switches on but does not heat up?
    • J
      Jessica BarreraAug 4, 2025
      If your Cristofoli Medical Equipment autoclave switches on but doesn't heat up, ensure you've pressed the start button after selecting the program. Also, if all the LEDs blink 3 times when pressing the start button, consult the section “How To Use The Vitale Class Autoclave”. If the problem persists, see “How To Proceed When Service Is Needed”. Another possible cause is incompatible voltage. In this case, when turning the equipment on, the LEDs Cancelado (Canceled), Aquecendo/Secando (Heating/Drying) and the LED that indicates capacity will be alight ( kgf/cm² for 127 V and 1,7 1,8 kgf/cm² for 220 V).
  • A
    Ann JohnsonAug 12, 2025
    Why does Cristofoli vitale 21 Medical Equipment autoclave take too long to build up pressure?
    • R
      Ricky MillerAug 12, 2025
      If the Cristofoli Medical Equipment autoclave takes too long to build up pressure, doesn't maintain pressure, and shows “CYCLE CANCELLED”, it could be due to several reasons. First, there might be voltage fluctuation, so consult your local energy provider. Second, check for a pressure or steam leak through the door gasket and perform “Preventive Maintenance”. Finally, ensure the chamber isn't overloaded, refer to “How to Use the Vitale Class CD Autoclave”.
  • M
    Michael BensonAug 19, 2025
    What to do if Cristofoli vitale 21 Medical Equipment autoclave pressure gets too high?
    • L
      Laura UnderwoodAug 19, 2025
      If your Cristofoli Medical Equipment autoclave pressure gets too high, activating the safety devices, the hose connected to the external steam exit may be obstructed. Remove the hose and clear any obstruction. Never use a common plastic hose. If the problem persists, consult “How To Proceed When Service Is Needed”.

Summary

Important Safety Information

Installation Instructions

Hydraulic Installation

Autoclave Components Identification

Safety Devices

Safety Notes

How to Use the Autoclave Vitale

Adverse Situations

Altitude Calibration

Technical Data

Preventive Maintenance

Preventive Maintenance Table

Troubleshooting

The Autoclave Does Not Switch On

Steps to diagnose and resolve issues when the autoclave fails to power on.

The Autoclave Pressure Gets Too High, Activating the Safety Devices

The Autoclave Takes Too Long to Build Up Pressure or Does Not Keep It, Indicating Cycle Cancelled

How to Proceed When Service is Needed

Warranty Terms

Cristofoli vitale 21 Specifications

General IconGeneral
BrandCristofoli
Modelvitale 21
CategoryMedical Equipment
LanguageEnglish