Specifications
5
-
2
INFORMATION
Cooling water temperature and flow rate should fall within the shaded area of Figure 5-1 to
maintain optimal performance. Operating outside of the shaded area may result in system
entering a warning or error state.
Figure 5-1: Cooling water requirement*
Figure 5-2: Cooling water pressure drop*
*When using a 50-50% mixture of ethylene glycol and water increase the flow rate 10%.
Pressure drop values will increase 40%.
CAUTION
The use of DI (De-Ionized) water is prohibited. DI water causes damage to the compressor package
and will not be covered under warranty.