Specifications
5
-
4
High-Pressure Atmospheric Relief Valve
The helium compressor package high-pressure atmospheric relief valve is set at
420 ± 5 psig (29 ± .34 bar). At pressures above 420 psig (29 bar) the relief valve will open
automatically and relieve pressure to the atmosphere.
Internal Motor Overload Switch
An internal motor overload switch, located inside the helium compressor package module, protects
the system by sensing excessive current draw and temperature. This switch automatically resets
itself after the helium compressor package module cools to an acceptable level.
Cold Head High Pressure Relief Valve
The cold head high pressure atmospheric relief valve is set at 425 ± 5 psig (29.3 ± .34 bar). At
pressures above 425 psig (29.3 bar) the valve will open automatically and relieve pressure to the
atmosphere.