Emissions Tester CCP700/800 Series Hints & Tips
10
High HC (Hydrocarbons)
Go to live readings and check HC, the reading should be less than 20 parts per million (PPM)
If no improvement, then remove right hand lter bowl. Check readings, if improved then clean
bowls, replace lters.
Remove probe see if it eventually drops to below 20. If it does, clean probe/hose, if that doesn't
help then replace hose/probe.
Oxygen Cell Failure
To check the Oxygen cell
Go into the Tools menu and select “Oxygen Cell Diagnostics” then run test.
The results should be between 5 and 13 if it’s anything else then the oxygen cell to be replaced.
If the customer has a replacement on site then they will need to replace the cell and then run
Install new Oxygen cell under: - Tools > Oxygen cell diagnostics> Install new Oxygen cell.
https://www.cryptontechnology.com/images/support/pdf/how-to-guides/o2-cell-expired.pdf
680 Wrong Date
If the unit is showing the wrong date this can be changed by
Petrol Menu > Tools and Setup > Service Mode > Engineer code > Set Time and Date
Changed the time and date accordingly, click continue and then exit back to the main menu
Do a live reading print out to make sure the date has been changed.
NB: Do not confuse this with calibration date being incorrect.
Emissions Tester CCP700/800 Series Hints & Tips
11
VTS printing
We have identied a small program error on a small number of Windows 10 machines running the
682 gas analyser. The issue can be identied by a lack of VTS information on the gas print out.
Should you experience this fault there is a simple solution as detailed below.
• Using WGTS3, enter the VTS details in the usual way.
• Close WGTS3.
• Launch Regedit (to do this type regedit after pressing the start button)
• Locate the below folder
[HKEY_CURRENT_USER\Software\Crypton\vts]
• Now highlight the vts folder and export the key (go to File > Export)
• Save the reg file to the desktop, call it VTS.reg
• Edit VTS.reg with a text editor (such as notepad) and change the key path in the file to:
• [HKEY_CURRENT_USER\Software\Crypton\WGTS3\vts]
Save the file, and then close it.
Double-click the reg file and accept the elevated prompt, then click "Yes" to the continue
message.
When this is complete then re-launch WGTS3 - the VTS details should appear on the printout.