Emissions Tester CCP700/800 Series Hints & Tips
12
Log File error
1) On the Gas program, select Service Mode on the right hand side
2) Get the calibration code off them
3) Give them the reply code
4) After this, customer should have lots more options down the right side
5) Select "Service Mode Diagnostics"
6) Then "Gas Analyser Serial Number Update"
7) Ensure that the serial number matches the one on the unit - customer MUST re-enter serial no.
Then, click on "Audit Log Diagnostics" (right hand side)
9) Will come up with a load of "gobble-dee-gook"
10) Just click Exit/Blue Arrow/X - back to the main screen
Smart Card Server
For systems that exhibit a ‘SmartCard Server’ issue in the UK Petrol test after the Diesel update has
been applied, they can be recovered with the following procedure.
In Control Panel, select WGTS3UK, then select Repair.
The following Windows Installer box will be displayed.
Type in the following path into the box
Emissions Tester CCP700/800 Series Hints & Tips
13
Then Click Ok.
This will restore the missing file.
If the Platform Configuration Utility has been run whilst the file is missing then use the Platform
Conguration Utility again to configure the Smartcard to an unused serial port.

Other manuals for CryptOn 700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CryptOn 700 Series and is the answer not in the manual?

CryptOn 700 Series Specifications

General IconGeneral
BrandCryptOn
Model700 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals