What to do if CU Medical Systems Medical Equipment i-PAD NF1200 does not turn on?
- AangelawatsonAug 1, 2025
If your CU Medical Systems i-PAD NF1200 doesn't turn on, there are several potential causes: * **Low Battery Charge:** Try a battery insertion test with a new battery pack. * **Discontinuity:** Check the battery pack contact spring J8 on the power board for damage and replace if necessary. * **Blown Fuse:** Inspect fuse F1 or F2 and replace it if it's blown, using only the manufacturer-recommended fuse. * **Power Switch Failure:** Check the ON/OFF switch SW2 on the main PCB board to ensure the poles are shorted when pressed; replace if faulty. * **Power Subsystem Fault:** Replace the main power board. If the issue continues, revert to the original board and contact the manufacturer.