Attach the Velcro strap.
SILENCE ALARM: return weight to
the sensor pad or press “RESET”.
See the Directions for Use if
operating in KEY MODE.
SILENCE ALARM: return weight to
the sensor pad or press “RESET”.
See Directions for Use if operating
in KEY MODE.
INCREASE DELAY: press and hold the “DELAY”
button (maximum setting 4 seconds for corded
and 5.5 seconds for wireless sensor pads).
DECREASE DELAY: press the “DELAY” button,
release, press again, and hold (minimum setting
0 seconds for corded and 1.5 seconds for
wireless pads).
Open the battery compartment:
1. Install batteries
2. If desired, set the “PAUSE” switch to “ON” (this
allows patient transfer for 30 seconds without
activating the alarm when pressing “RESET”
button).
3.
Close the cover and secure the screw
NOTE:
Pad life can be displayed only when
PAUSE is o.
TONE SETTING: press and release the “TONE”
button to select one of four tone options.
1. Press and hold the black programming button on the back of the monitor,
until the monitor beeps three times and displays “PROGRAM MODE”
2. Firmly place pressure on pad and release. A long beep confirms success
and the monitor will display “PAD SYNC'ED”.
1. Press and hold the black programming button on the back of the monitor,
until the monitor beeps three times and displays “PROGRAM MODE”
2. Firmly place pressure on pad and release. A long beep confirms success
and the monitor will display “PAD 2 SYNC'ED”.
NORMAL OPERATION: Key Mode switch in up
position
TAMPER-RESISTANT OPERATION:
Key Mode
switch in down position
requires use of the
override key to be placed over “KEY” icon on face
of monitor in order to silence alarm and change
volume, delay and tone (see the Directions for
Use for full instructions).
INCREASE VOLUME: Press and hold the “VOL
button (maximum volume is 7 bars).
DECREASE VOLUME: Press the “VOL” button,
release, press again and hold (minimum is
volume o)
NOTE:
Volume can be turned o when a nurse call
connection cable is connected.
TEST ALARM: Put weight on the
sensor pad until you hear a double
beep, “PAD IN USE” is displayed,
and the green LED blinks. Release
weight to activate alarm.
TEST ALARM: Put weight on the
sensor pad until you hear a double
beep, “PAD 2 IN USE” is displayed,
and the green LED blinks. Release
weight to activate alarm.
1
1
1
3
3
5
2
6 7
4
2
2
MONITOR SETUP
SYNCHRONIZING ONE CORDLESS SENSOR
(refer to transmitter setup)
SYNCHRONIZING A SECOND CORDLESS SENSOR
©2016 Curbell Medical Products, Inc. MAP1607A
www.curbellmedical.com • 1-888-222-2188
CSM-BC400 QUICK SETUP GUIDE
NOTE: Consult the “Directions For Use” included with monitor for
complete instructions, troubleshooting, and safety information
Question and Answer IconNeed help?

Do you have a question about the Curbell CSM-BC400 and is the answer not in the manual?

Curbell CSM-BC400 Specifications

General IconGeneral
BrandCurbell
ModelCSM-BC400
CategoryControl Unit
LanguageEnglish

Summary

Monitor Setup and Sensor Synchronization

Volume and Delay Adjustments

Configure audio alerts and alarm delay times for the monitor.

Tone and Operation Mode Selection

Choose alert tones and switch between normal and tamper-resistant modes.

Single Cordless Sensor Synchronization

Pair a single wireless sensor pad with the CSM-BC400 monitor.

Second Cordless Sensor Synchronization

Pair a second wireless sensor pad with the CSM-BC400 monitor.

Alarm Testing and Silencing

Verify sensor functionality and how to silence active alarms.

Unsynchronizing and Corded Sensor Use

Unsynchronizing Monitor and Pad

Procedure to disconnect a sensor pad from the monitor.

Corded Sensor Pad Connection and Use

How to connect a corded sensor and test its alarm function.

System Setup and Connectivity

System Setup and Pad Placement

Steps for securing the monitor and positioning the sensor pad.

Nurse Call and Call Cord Connection

Instructions for connecting to nurse call systems and using call cords.

Troubleshooting Common Issues

No Power or Indicator Light Issues

Diagnose problems with the monitor not powering on or the 'in use' light.

Low Battery or Replace Battery Display

Resolving issues related to battery status notifications.

Wireless Sensor Synchronization Problems

Troubleshoot why a wireless sensor pad is not connecting to the monitor.

Alarm Reset and Nurse Call System Issues

Address problems with alarm reset functionality and nurse call alerts.

False Alarms and Corded Sensor Issues

Resolve false alarms, pad placement, and corded sensor connection problems.

Lost Signal Alert Troubleshooting

Range and Transmitter Placement

Ensure proper sensor pad and transmitter positioning for optimal range.

Signal Interference and Best Practices

Identify factors causing signal loss and recommended setup for performance.