Page #30 background imageLoading...
Page #30 background image
Electrical
fast
transient/
burst IEC
61000-4-4
±2kV for
power supply
lines
±1kV for
input/output
lines
Not applicable
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC
61000-4-5
±1kV
differential
mode
±2kV common
mode
Not applicable Mains power quality
should be that of a
typical commercial
or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
<5% UT
(>95% dip in
UT) for 0.5
cycle
40% UT
(60% dip in
UT) for 5
Cycles
70% UT
(30% dip in
UT) for 25
Cycles
<5% UT
(>95% dip in
UT) for 5 sec
Not applicable Mains power quality
should be that of a
typical commercial or
hospital environment.
Power
frequency
(50/60Hz)
magnetic
field IEC
61000-4-8
3 A/m 3 A/m Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE UT is the a.c mains voltage prior to application of the test level.
30

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Current Solutions LG SMART TENS and is the answer not in the manual?

Current Solutions LG SMART TENS Specifications

General IconGeneral
BrandCurrent Solutions
ModelLG SMART TENS
CategoryPersonal Care Products
LanguageEnglish

Related product manuals