To measure accurately, please degauss and zero set the probe, and make sure the probe is locked during
the process.
Set the input impedance of the oscilloscope to 1MΩ(default)
Make sure the probe is locked during testing.
The probe should be away from the interference source like transformer. The method to judge if the
probe is interfered is to put the probe close to circuit under test. IF there’s any output, there could be
interference in the testing environment because the probe is not on the circuit yet.
The current under test should not surpass the limit value of the probe.
Please always maintain your probe and do not use it in the humid environment
If there’s anything wrong with the probe, please set it back for repairing. If you dismantled the device
on your own, we won’t guarantee for repairing.
10. Packing list
Packing list
ITEM
Quantity
Probe
1
DC-12V/1A adapter
1
BNC connecting line
1
Instruction manual
1
Guarantee card
1
Test report
1
©Zhiyong Electronics, 2023
Published in China, Step. 1, 2023
CYBERTEK
TESTLINK
TESTLINK Corporation
Addr(14348) B 709 (XI Tower) , 67 Saebitgongwon-ro, Gwangmyeong-si, Gyeonggi-do, Korea
Tel: +82.2.6299.7588
1588-8679
Fax: +82.2.6299.7581
Email: sales@testlink.co.kr
URL: http://www.testlink.co.kr

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cybertek HCP8000 Series and is the answer not in the manual?

Cybertek HCP8000 Series Specifications

General IconGeneral
BrandCybertek
ModelHCP8000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals