5-3. Simulation Calibration Mode (Without Test Weight) – Mode 2.
- This calibration Method will be useful to make calibration more than 10ton capacity setting.
- Guaranteed resolution will be 1/5,000 and if you need higher resolution, please make
calibration with Test weight.
Step 1. Enter to the “SET-CAL” mode
Please turn on the Power + with pressing display
Press key display.
Press key to start “Simulation Calibration Mode”.
Remarks : Go to next step with save key / Back to previous step key Step 2.
Step 2. Digit (Division)setting ( display)
Press key and stop the optimal Digit value.
(Whenever pressing key, digit value will be changed, like 01-02-05-10-20-50)
Press key and save change and move to next step.
Step 3. Max. Capacity of Load cell ( display)
Input Max. Capacity of Scale with No. keys.
- Under this step, input Total sum of each load cell’s Max. Capacity. (Not weighing Scale)
- The Max. Capacity of load cell is stated on “Test report” or “Label”.
- If you installed 4 load cells, and each load cell’s Max. Capacity is 500kg, then you have to input
2,000kg, as a Max. Capacity.
Input Capacity and press key, and move to next step.