Step 4. Zero Balance setting ( display)
Make empty the scale part, and press key.
Indicator check the current Zero balance and save the value and move next step.
Step 5. Span value calculation ( )
Input prepared Test weight value with No. keys.
And press key.
Then, display will show and then, load prepared test weight unit on the scale.
After a few seconds(to remove the vibration effect), press key.
Then, indicator will calculate Span value and move the next step.
Caution
For the precise Span calibration, please prepare Test weight unit, at least 10% of Max. capacity
of Scale.
Step 6. Check Span Value and Finish the Calibration( )
Check the Calculated Span value.
And after 3sec, will displayed automatically and move to weighing Mode.
C-END