Datex-Ohmeda Compact Airway modules
S/5
TM
Compact Airway Module, M-CAiOVX (rev. 01)
S/5
TM
Compact Airway Module, M-CAiOV (rev. 03)
S/5
TM
Compact Airway Module, M-CAiO (rev. 02)
S/5
TM
Compact Airway Module, M-COVX (rev. 02)
S/5
TM
Compact Airway Module, M-COV (rev. 03)
S/5
TM
Compact Airway Module, M-CO (rev. 02)
S/5
TM
Compact Airway Module, M-C (rev. 02)
Technical Reference Manual Slot
All specifications are subject to change without notice.
Document No. 800 1009-1
June 2001
Datex-Ohmeda Inc.
3030 Ohmeda Drive
53707-7550 MADISON, WIS
USA
Tel. +1-608-221 1551, Fax +1-608-222 9147
www.us.datex-ohmeda.com
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

Table of Contents

Questions and Answers

  • K
    Kathleen RodriguezAug 18, 2025
    Why does my Datex-Ohmeda Test Equipment show a REBREATHING or FiCO high message?
    • K
      Karla EricksonAug 18, 2025
      If your Datex-Ohmeda Test Equipment displays the message 'REBREATHING' or 'FiCO high', it indicates that the CO concentration in the inspiratory air is too high. This is possibly due to a saturated CO absorber in the ventilation system. To resolve this, change the ventilation absorber.
  • S
    Suzanne SantiagoAug 20, 2025
    What to do if my Datex-Ohmeda Test Equipment displays 'SAMPLE LINE BLOCKED' and 'Air leak' message?
    • D
      Dr. Justin Mitchell DDSAug 20, 2025
      If your Datex-Ohmeda Test Equipment shows a 'SAMPLE LINE BLOCKED' message accompanied by an 'Air leak' message, it means the sampling line or water trap is blocked, or the water trap container is full. If the occlusion persists, check the internal tubing for blockages.
  • S
    Scott EvansAug 23, 2025
    What does 'REPLACE D-FEND' message mean on Datex-Ohmeda Test Equipment?
    • K
      Kelly LeeAug 23, 2025
      If your Datex-Ohmeda Test Equipment displays a 'REPLACE D-FEND' message (or 'Replace water trap' message), it indicates residue build-up on the water trap membrane, decreasing air flow. Replace the D-fend to resolve this issue.
  • H
    Henry SalasAug 26, 2025
    How to fix SENSOR INOP message on Datex-Ohmeda Test Equipment?
    • M
      marquezandreaAug 26, 2025
      If your Datex-Ohmeda Test Equipment is displaying 'SENSOR INOP' message, it could be due to high temperature or a communication error. Check the fan and filter at the front panel. Also, check the timeout and bad checksum values in the service menu.
  • N
    Nicole DawsonAug 27, 2025
    What causes 'xx ZEROING ERROR' on Datex-Ohmeda Test Equipment?
    • S
      Shannon OwensAug 27, 2025
      If your Datex-Ohmeda Test Equipment shows a 'xx ZEROING ERROR' message, it means condensation or residual gases are affecting the zero measurement. Allow the module to run drawing room air for half an hour, then calibrate again.
  • K
    Karen Pitts MDAug 30, 2025
    What does MVexp
    • S
      sherry03Aug 30, 2025
      If your Datex-Ohmeda Test Equipment shows a 'MVexp
  • K
    Kenneth VelasquezSep 1, 2025
    What does (Disconnection) (MVexp < 0.5 l/min message) mean on Datex-Ohmeda Datex-Ohmeda Test Equipment?
    • J
      James NicholsSep 1, 2025
      If your Datex-Ohmeda Test Equipment displays '(Disconnection) (MVexp < 0.5 l/min message) Low volumes', it means tidal volumes are too small for accurate measurement (this is not shown during Apnea), and gas sampling is working correctly. Check the D-lite connections and D-lite tubing.
  • Y
    yrichardsSep 3, 2025
    What does inspiratory TV>exp TV mean on Datex-Ohmeda Test Equipment?
    • G
      Gary Moore DDSSep 3, 2025
      If your Datex-Ohmeda Test Equipment displays 'inspiratory TV>exp TV', it indicates a leak in the lungs or an ET tube cuff leak. Check for leakages and perform a leak test. You may also need to change the tubing and D-lite. Avoid active humidification, and ensure gas sampling is connected only and always to D-lite. Check the D-fend as well.
  • C
    Christopher KochSep 6, 2025
    What does loop overscale monitored volumes < set volumes mean on Datex-Ohmeda Test Equipment?
    • M
      Molly WeeksSep 6, 2025
      If your Datex-Ohmeda Test Equipment displays 'loop overscale monitored volumes < set volumes', it means the wrong scale is selected or there's a leak between the ventilator and D-lite. Change the scaling and check the ventilator connections.
  • M
    Michael CameronSep 8, 2025
    What does exp TV> inspiratory TV mean on Datex-Ohmeda Test Equipment?
    • E
      Edward RiveraSep 8, 2025
      If your Datex-Ohmeda Test Equipment displays 'exp TV> inspiratory TV', it indicates a spirometry tube leak. Check for leakages and perform a leak test.

Summary

INTRODUCTION

About M-CAIOVX;SERVICE module

Details about the M-CAIOVX/SERVICE module for service purposes, including differences from standard models.

1 SPECIFICATIONS

1.1 General specifications

Provides physical dimensions, weight, operating temperature, power consumption, and electrical safety details.

1.2 Typical performance

Lists measurement ranges and accuracy for CO2, O2, and N2O parameters.

1.3 Gas specifications

Outlines conditions for airway humidity, sampling rate, response times, and warm-up periods.

1.4 Patient spirometry specifications

Details parameters like airway pressure, flow, tidal volume, minute volume, compliance, and resistance.

1.5 Gas exchange specifications

Covers VO2, VCO2, and RQ measurements, including range, resolution, and accuracy.

2 FUNCTIONAL DESCRIPTION

2.1 Measurement principle

Explains the measurement principles for CO2, N2O, agent, O2, and spirometry.

2.2 Main components

Lists the primary components of the compact airway modules, such as TPX, OM, and PVX units.

2.3 Connectors and signals

Describes the module bus connector pin layout and signal descriptions for interface.

3 SERVICE PROCEDURES

3.1 General service information

Provides guidelines for field service, safety precautions, and factory repair limitations.

3.2 Service check

Outlines the procedures for performing a service check on the module.

3.3 Disassembly and reassembly

Details the step-by-step process for disassembling and reassembling the module.

3.4 Adjustments and calibrations

Covers adjustments and calibrations for gas sampling, flow rates, and gas calibration.

4 TROUBLESHOOTING

4.1 Troubleshooting charts

Provides tables linking problems to possible causes and treatments for various functions.

4.2 Gas sampling system troubleshooting

Addresses issues related to leaks or blockages in the gas sampling system.

4.3 OM measuring unit troubleshooting

Discusses troubleshooting steps and cautions for the Oxygen (OM) measuring unit.

4.4 TPX measuring unit troubleshooting

Provides guidance and factory repair information for the TPX measuring unit.

4.5 PVX measuring unit troubleshooting

Explains troubleshooting for the PVX measuring unit and replacement of the unit.

4.6 CPU board troubleshooting

Offers basic checks for the CPU board and its installed ICs.

4.7 Error messages

Lists common error messages and their explanations for diagnosing issues.

5 SERVICE MENU

5.1 General menu

Explains the General Service Menu, module configuration, and service data fields.

5.2 Gases service menu

Details the parameters and options available within the Gases service menu.

5.3 Spirometry service menu

Covers the Spirometry service menu, including sensor type, pressure, and flow adjustments.

6 SPARE PARTS

6.1 Spare parts list

Lists part numbers for various components of the compact airway modules.

7 EARLIER REVISIONS

APPENDIX A

Service check form, S;5 Compact Airway modules

Refers to the service check form for recording maintenance and testing procedures.