Datex-Ohmeda S/5 M-CAiOVX - User Manual

Datex-Ohmeda S/5 M-CAiOVX
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Datex-Ohmeda Compact Airway modules
S/5
TM
Compact Airway Module, M-CAiOVX (rev. 01)
S/5
TM
Compact Airway Module, M-CAiOV (rev. 03)
S/5
TM
Compact Airway Module, M-CAiO (rev. 02)
S/5
TM
Compact Airway Module, M-COVX (rev. 02)
S/5
TM
Compact Airway Module, M-COV (rev. 03)
S/5
TM
Compact Airway Module, M-CO (rev. 02)
S/5
TM
Compact Airway Module, M-C (rev. 02)
Technical Reference Manual Slot
All specifications are subject to change without notice.
Document No. 800 1009-1
June 2001
Datex-Ohmeda Inc.
3030 Ohmeda Drive
53707-7550 MADISON, WIS
USA
Tel. +1-608-221 1551, Fax +1-608-222 9147
www.us.datex-ohmeda.com
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

Table of Contents

Questions and Answers

  • A
    Alyssa BurnsAug 19, 2025
    Why does my Datex-Ohmeda Test Equipment show a REBREATHING or FiCO high message?
    • M
      Michael KaiserAug 19, 2025
      If your Datex-Ohmeda Test Equipment displays the message 'REBREATHING' or 'FiCO high', it indicates that the CO concentration in the inspiratory air is too high. This is possibly due to a saturated CO absorber in the ventilation system. To resolve this, change the ventilation absorber.
  • T
    Tracy LongAug 21, 2025
    What to do if my Datex-Ohmeda Test Equipment displays 'SAMPLE LINE BLOCKED' and 'Air leak' message?
    • A
      ajohnsonAug 21, 2025
      If your Datex-Ohmeda Test Equipment shows a 'SAMPLE LINE BLOCKED' message accompanied by an 'Air leak' message, it means the sampling line or water trap is blocked, or the water trap container is full. If the occlusion persists, check the internal tubing for blockages.
  • S
    Scott DuncanAug 23, 2025
    What does 'REPLACE D-FEND' message mean on Datex-Ohmeda Test Equipment?
    • J
      Jacob SheltonAug 25, 2025
      If your Datex-Ohmeda Test Equipment displays a 'REPLACE D-FEND' message (or 'Replace water trap' message), it indicates residue build-up on the water trap membrane, decreasing air flow. Replace the D-fend to resolve this issue.
  • T
    Tyler RichardsAug 26, 2025
    How to fix SENSOR INOP message on Datex-Ohmeda Test Equipment?
    • A
      Amanda MossAug 26, 2025
      If your Datex-Ohmeda Test Equipment is displaying 'SENSOR INOP' message, it could be due to high temperature or a communication error. Check the fan and filter at the front panel. Also, check the timeout and bad checksum values in the service menu.
  • R
    Ricky MartinezAug 28, 2025
    What causes 'xx ZEROING ERROR' on Datex-Ohmeda Test Equipment?
    • T
      tiffany42Aug 28, 2025
      If your Datex-Ohmeda Test Equipment shows a 'xx ZEROING ERROR' message, it means condensation or residual gases are affecting the zero measurement. Allow the module to run drawing room air for half an hour, then calibrate again.
  • H
    heatherrayAug 30, 2025
    What does MVexp
    • B
      Brenda BishopAug 30, 2025
      If your Datex-Ohmeda Test Equipment shows a 'MVexp
  • L
    lanthonySep 1, 2025
    What does (Disconnection) (MVexp < 0.5 l/min message) mean on Datex-Ohmeda Datex-Ohmeda Test Equipment?
    • J
      Jennifer NewtonSep 1, 2025
      If your Datex-Ohmeda Test Equipment displays '(Disconnection) (MVexp < 0.5 l/min message) Low volumes', it means tidal volumes are too small for accurate measurement (this is not shown during Apnea), and gas sampling is working correctly. Check the D-lite connections and D-lite tubing.
  • H
    Heather ProctorSep 3, 2025
    What does inspiratory TV>exp TV mean on Datex-Ohmeda Test Equipment?
    • D
      David MillerSep 4, 2025
      If your Datex-Ohmeda Test Equipment displays 'inspiratory TV>exp TV', it indicates a leak in the lungs or an ET tube cuff leak. Check for leakages and perform a leak test. You may also need to change the tubing and D-lite. Avoid active humidification, and ensure gas sampling is connected only and always to D-lite. Check the D-fend as well.
  • W
    washingtonheatherSep 6, 2025
    What does loop overscale monitored volumes < set volumes mean on Datex-Ohmeda Test Equipment?
    • K
      Kimberly BrownSep 6, 2025
      If your Datex-Ohmeda Test Equipment displays 'loop overscale monitored volumes < set volumes', it means the wrong scale is selected or there's a leak between the ventilator and D-lite. Change the scaling and check the ventilator connections.
  • J
    Jill ClarkSep 8, 2025
    What does exp TV> inspiratory TV mean on Datex-Ohmeda Test Equipment?
    • D
      Darren SmithSep 8, 2025
      If your Datex-Ohmeda Test Equipment displays 'exp TV> inspiratory TV', it indicates a spirometry tube leak. Check for leakages and perform a leak test.

Summary

INTRODUCTION

About M-CAIOVX;SERVICE Module

Details the purpose, specifications, and physical differences of the M-CAIOVX/SERVICE module for service use.

SPECIFICATIONS

General Specifications

Covers module size, weight, operating/storage temperatures, pressure, humidity, power consumption, and electrical shock protection.

Typical Performance Metrics

Specifies performance metrics for CO2, O2, N2O, respiration rate, anesthetic agents, and MAC.

Gas Specifications

Outlines parameters like airway humidity, sampling rate, delays, and warm-up times under normal and exceptional conditions.

Patient Spirometry Specifications

Details specifications for airway pressure, flow, tidal volume, minute volume, compliance, and resistance.

Gas Exchange Specifications

Covers VO2, VCO2, and RQ measurements, including range, resolution, and accuracy.

FUNCTIONAL DESCRIPTION

Measurement Principles

Explains the principles behind CO2, N2O, agent, O2, spirometry, and gas exchange measurements.

Main Components Overview

Lists and describes the primary components of the compact airway modules, including boards and units.

Connectors and Signals

Details the pin layout and signal descriptions for the module bus connector.

SERVICE PROCEDURES

General Service Information

Provides guidelines for field service, factory repairs, and personnel qualifications.

Service Check Procedures

Outlines the procedures for performing a service check, including recommended tools and parts.

Disassembly and Reassembly

Provides step-by-step instructions for taking apart and putting back together module components.

Adjustments and Calibrations

Covers procedures for gas sampling system adjustment, flow rate measurement, and calibration.

TROUBLESHOOTING

Troubleshooting Charts

Lists common problems and their possible causes and treatments for various measurements.

Gas Sampling System Troubleshooting

Addresses issues related to leaks, blockages, pump failures, and flow rate problems in the sampling system.

Module Unit Troubleshooting

Details troubleshooting steps for OM, TPX, PVX units, and the CPU board.

Error Messages

Explains various error messages displayed by the module and their meanings.

SERVICE MENU

Service Menu - General

Describes the general service menu, including module configuration and service data fields.

Service Menu - Gases

Details options for noise measurement, gain adjustment, fall time, and pump/valve controls.

Service Menu - Spirometry

Covers adjustments and data display related to spirometry measurements like pressure, flow, and volume.

EARLIER REVISIONS

APPENDIX A

Appendix A: Service Check Form

A form for documenting service checks, including pass/fail criteria for various module components.

SPARE PARTS

Spare Parts List

Provides lists of order numbers for various module components and accessories.

Datex-Ohmeda S/5 M-CAiOVX Specifications

General IconGeneral
BrandDatex-Ohmeda
ModelS/5 M-CAiOVX
CategoryTest Equipment
LanguageEnglish