Why is LAI output 0 on my Decagon Devices Measuring Instruments?
- CCarrie BradySep 12, 2025
If your LAI output is zero despite valid PAR readings above and below the canopy on your Decagon Devices Measuring Instruments, it's likely that the time or latitude/longitude settings are incorrect, causing Z to exceed 90 degrees. To fix this: 1. Go to the set date/time menu. 2. Ensure the time is in 24-hour format. 3. Verify the correct sign for your latitude and longitude.