5
DAC-530E-EN-BH
2 Dangers, Warnings and Cautions
This chapter includes a list of danger, warning, and caution messages that relate to the Defibtech
DDU-120 AED and its accessories. Many of these messages are repeated elsewhere in this
User Manual and on the DDU-120 AED or accessories. The entire list is presented here for
convenience.
DANGER: Immediate hazards that will result in serious personal injury
or death.
WARNING: Conditions, hazards, or unsafe practices that may result in
serious personal injury or death.
CAUTION: Conditions, hazards, or unsafe practices that may result in
minor personal injury, damage to the DDU-120 AED, or loss
of data.
2.1 Shock, Fire Hazard, Explosion
2.1.1 Electricity
DANGER
Hazardous electrical output. This equipment is for use only by qualified
personnel.
2.1.2 Battery Pack
CAUTION
Follow all battery pack labeling instructions. Do not install battery packs after
the expiration date.
WARNING
Lithium battery packs are not rechargeable. Any attempt to recharge a lithium
battery pack may result in fire or explosion.
WARNING
Do not immerse battery pack in water or other liquids. Immersion in fluids
may result in fire or explosion.