ElastiMeter User Manual version 1.6
15.8.2018
11
TECHNICAL DATA
System components
o ElastiMeter measuring unit with integrated probe
o Batteries (installed)
o User manual (USB flash drive + printed)
o Carrying case
Accessories
o DMC data collection software (USB flash drive)
o Wireless receiver unit
Power supply
o 2 x 1.5 V AA alkaline battery (IEC: type LR6)
o Power consumption: 360 mW 495 mW (Wireless mode off on)
ElastiMeter measuring unit
o Dimensions: 189 x 45 x 35 mm
o Weight with batteries: 153 g
o Protection class IP52
o Measurement completion time (ISE): 0.5 seconds
o Measurement repeat cycle (ISE) 10 measurements per minute typically
o Measured parameter: Instant Skin Elasticity (ISE)
o Measuring SI unit: N/m (Newton per meter)
o Operating range: 10 160 N/m
o Force sensor accuracy: ±3 %
o Coefficient variation (SD/mean) measured on normal skin
typically < 10 %
Integrated probe
o Indenter length: 0.6 mm
o Indenter diameter: 2.5 mm
o Indenter type: stainless steel (AISI 316L) with rounded tip
o Base plate diameter: 23 mm
o Base plate material: PTFE
o Maximum force applied to either the base plate or the indenter: 10 N
*