UNIT SET UP
UNPACKING
Verify that all items are accounted for and undamaged:
Veterinary Scaler unit with integral foot pedal
Ultrasonic Scaler hand piece
TFI#3 Scaling Insert
Quick Release Valve water connection
Water Line with Filter installed
Water Supply
Install the ¼” male iron pipe thread coupler body into a source of cold drinkable water with a
pressure of no less than 20 PSI and no more than 50 PSI. If pressure exceeds 50 PSI a regulator
will be required. It is necessary to have a shut off valve in series with the coupler body. The
internal water stop in the coupler body is not meant to act in lieu of a shutoff.
CAUTION! The water itself must conform to CDC standards in order to be free of potential
bacterial contamination and be relatively soft. Excess mineral content will ultimately degrade the
water system in the 2000 Plus unit, requiring repair. All plumbing work must comply with local
building and plumbing codes. It is strongly advised that a licensed plumber be employed to install
the coupling body and shutoff valve into the building water system.
Flush the newly installed quick release coupler via the water hose thoroughly before connecting
to the filter in order to clear out any debris remaining from the plumbing work. Connect the
water supply hose going into the unit to the filter via the Luer Lock fitting. Connect the water line
to the filter via the Luer Lock and then into water supply through the quick disconnect coupling
using the thumb lever release. IMPORTANT! REPLACE THE FILTER EVERY 30 DAYS.
NOTE! Although the flushing instruction given above is meant for one time use on new
installations, it should be repeated if any repair work is done to the building plumbing system or if
building water pressure is lost. Verify that cold water is present after building plumbing work.