Chapter 12 Descriptions of Parameter SettingsC2000 Plus 
 
12.1-10-14
 
10-41 
I/F Mode, Id Current Low Pass-Filter Time 
 
 
   
 Default: 0.2 
 
 
Settings 0.0–6.0 sec. 
  Set the filter time for Pr.10-31. Smoothly increases the magnetic field to the current command 
setting value under the I/F mode. 
  If you want to slowly increase the size of Id, increase the filter time to avoid a step phenomenon 
occurs when starting current output. When decrease the filter time (minimum value is 0), the 
current rises faster, then a step phenomenon occurs. 
 
10-42 
Initial Angle Detection Pulse Value 
 
 
   
 Default: 1.0 
 
 
Settings 0.0–3.0 
  The angle detection is fixed to 3: Use the pulse injection method to start. The parameter 
influences the value of the pulse during the angle detection. The larger the pulse, the higher the 
accuracy of rotator’s position. A larger pulse might cause oc. 
  Increase the parameter when the running direction and the command are opposite during start-
up. If oc occurs at start-up, then decrease the parameter. 
  Refer to Section 12-2 Adjustment & Application for detailed motor adjustment procedure. 
 
10-43 
PG Card Version 
 
 
   
  Default: Read only 
 
 
Settings 0.00–655.35 
 Corresponding versions for reference: 
PG02U  21.XX 
PG01U  31.XX 
PG01O / PG01L  11.XX 
PG02O / PG02L  14.XX 
PG01R  41.XX 
 
 
10-47 
PG1 Pulse Imputation Scaling Factor 
 
 
   
 Default: 0 
 
 
Settings 
0: x1 
1: x2 
2: x4 
3: x8 
 Use Pr.10-47 to set interpolation magnification of the PG1 Sin/Cos signal. After the interpolation 
is finished, the encoder PPR (Pulses per Revolution) = Pr.10-01 ×2
Pr.10-47
× 4. The larger the 
interpolation magnification, the more accurate the positioning. 
 Example: 
When Pr.10-01 = 128 and Pr.10-47 = 0, PPR= 128 × 2
0
 × 4 (four-time frequency) = 1024. 
When Pr.10-01 = 128 and Pr.10-47 = 3, PPR= 128 × 2
3
 × 4 (four-time frequency) = 8192. 
 
10-49 
Zero Voltage Time during Start-up 
 
 
   
 Default: 0.000 
 
 
Settings 0.000–60.000 sec.