Professional LED Light System for All of Surgery Application
6
Electromagnetic waves discharged from MRI can cause malfunction of this product.
Please keep away any device that can influence this product.
The LIGHTHEAD must be connected to a backup power supply(BATTERY PACK or
UPS).
Always should be checked Battery capacity level before use by pressing the BATTERY
CHECK BUTTON. A backup lifetime of a BATTERY PACK or UPS should be greater than
3 hours on power cut condition
DENTIS is not responsible for the customer’s backup power supply.(BATTERY PACK
or UPS)
Take care when handling the circuit boards: these boards are supplied in an
electrostatic envelope and must be handled with great care.
Do not look directly into light source(LED).
The LIGHTHEAD brakes are adjusted during installation. Like all mechanical parts,
the brakes are subject to wear.
Read just the brakes if the LIGHTHEAD no longer remains steady in any position.
Check the condition of the mounting surface.
The operation and safety of the device may be affected by the removal of certain
components during servicing operations.