Installation and handover report
Estimated amount of time required: approx. 45 min
Please check all connections, the device's oil and water supply, and the menu settings in advance.
Device data
Designation of device: DAC Universal
Device type: Cleaning and disinfection device as per EN ISO 15883-1/-2
Location:
Serial no.:
Manufacturer: Sirona Dental Systems GmbH
Installation date:
Year of manufacture:
Medium Target Actual
Conductivity of the water < 3 µS/cm in routine operation
Flow pressure Manufacturer's specification:
5 - 8 bar
Compressed air – measured flow pressure under real conditions
60 standard liters per minute
Minimum flow rate with fixed water connection 200 ml/min.
Presentation of the DAC Universal
Presentation of the scope of functions of the DAC Universal (cleaning, care, and disinfection) and the requirements for
the location and the environmental conditions.
☐
Explanation of the user interface and menu navigation.
☐
View of the instructions for use and explanation of the structure and contents of the instructions for use.
☐
Presentation of the contents of the USB stick DAC Universal (Touch).
☐
Explanation of the Check & Clean maintenance workflow (200 & 50) using the Check & Clean process within the soft-
ware.
Note on the necessity for and intervals of the Check & Clean process
☐
Explanation of the function key "Check & Clean 50" or "Check & Clean 200" of the start screen.
☐
Release notes
Successful cleaning must be rechecked visually after removing the instruments prior to release.
☐
Before use (on a patient), the spray function on the instrument must be tested.
☐
If reprocessing with the DAC Universal was interrupted by the user or the device itself, the process must be repeated,
as it was not successfully completed.
☐
Documentation
Note the operator's documentation requirement.
Presentation of the documentation options
☐