Dentsply Sirona 6Operation
Operating Instructions inFire HTC speed 6.2Switching on the unit
båÖäáëÜ
Description of the columns on the left side with the freely programmed
furnace programs
The laboratory furnace heats up and cools down in up to four temperature
steps. It is possible to heat or cool down within the program run.
6.2
Switching on the unit
Switch-on
✔ The unit has been properly set up and connected to a power outlet.
➢ Turn on the main switch.
The main switch is illuminated and the LCD screen shows a status
display for 3 seconds before the standby display is shown.
The furnace door opens automatically after a cooled furnace has been
switched on.
Status display
6.3
Practical use
Explanation The furnace can be used immediately after entering the required
program.
Column Designation Description
Heating rate (heating and
cooling down speed)
The heating and cooling down speed of the corresponding level is
set up / displayed in this column. The heating and cooling down
speed can range from 1 °C / min (33.8 °F / min) to 99 °C / min (210
°F / min).
Holding temperature The temperature of the corresponding level is set / displayed in this
column.
Holding time The holding time of the corresponding step is set / displayed in this
column.
Line 1 (S4) Model number (model no.:)
Line 2 (S3) Order number (REF:)
Line 3 (S2) Serial number of electronics (serial no.:)
Line 4 (S1) Hardware version (HW:) and software version (SW:)
NOTICE
Risk of breakage
The sintering trays to be used are fragile. They break with strong
temperature fluctuations and under mechanical stress.