61 30 459 D 3497
D 3497.201.01.19.02 09.2017
33
Dentsply Sirona 6Operation
Operating Instructions inFire HTC speed 6.5Non-precious metal (NPM) sintering (optional)
båÖäáëÜ
Determining the filling level of the compressed gas cylinder:
The filling level of the compressed gas cylinder can be ascertained from
the manometer. A new argon compressed gas cylinder is filled with 200
bar. If the pressure displayed is only half of the pressure, i.e. 100 bar, then
the cylinder is half empty.
Example:
a 50-liter compressed gas cylinder of argon with a filling pressure of 200
bar contains approx. 10,000 liters of argon.
The consumption per inCoris CC sintering cycle is approx. 270 liters at a
flow rate of 1.1 l / min.
As a result approx. 35 inCoris CC sintering cycles are possible with a 50-
liter compressed gas cylinder of argon.
6.5.5 Sintering pre-sintered NPMs (CoCr-base)
(not supplied depending on the standard equipment)
The set for the sintering of inCoris CC/CCB or “Crypton for inLab” consists
of the following components:
inCoris CCB set
A Furnace door with gas
socket for inCoris CC/CCB
D Base plate made of
high-performance ceramics
B Sintering tray made of
high-performance ceramics
E Sintering bell made of
high-performance ceramics
C Cover for sintering tray
made of high-performance
ceramics
F Zirconium oxide beads
Ø = 1.25 mm