EN - 6 A-4650D
COMPRESSOR CLEANING
1. With power switch in the “Off” position, unplug power cord from wall outlet.
2. Wipe outside of the compressor cabinet with a clean damp cloth every few days to keep dust-free.
DANGER
Do not submerge in water; doing so will result in compressor damage.
FILTER CHANGE
1. Filtershouldbechangedevery6monthsorsooneriflterturnscompletelygrayincolor.
2. Removeltercapbygraspingitrmlyandpullingoutoftheunit
3. Removedirtylterwithngersanddiscard.
4. ReplacewithanewDeVilbisslter(part#5650D-602).AdditionalltersshouldbepurchasedfromyourDeVilbiss
provider.
5. Pushltercapwithnewlterintoposition.
CAUTION– Do not reuse or wash lter or substitute any other material such as cotton for a DeVilbiss air-inlet lter. It will
result in compressor damage.
NOTE– An extra lter is packaged with the 4650 series units.
MAINTENANCE
AllmaintenancemustbeperformedbyaqualiedDeVilbissproviderorauthorizedservicecenter.
DANGER
Electric shock hazard. Do not remove compressor cabinet. All disassembly and maintenance must be done by a
qualied DeVilbiss provider.
SPECIFICATIONS
Size......................................................... 7” W x 3.8” H x 13” D (18 cm x 10 cm x 33 cm)
Weight ...............................................................................5.3 lbs. (2.40 kg)
Electrical Requirements ....................................................2.5 amps, 115 VAC (±10%) 60 Hz
Sound Level ...............................................................................53-57 dBA
Power Consumption..................................................................140 watts maximum
Maximum Compressor Pressure .........................................................35 psig (241 kPa)*
Compressor Free Air Flow .......................................................................13 lpm*
Operating Temperature Range .................................................+40° to +104°F (+5 to +40°C)
Operating Humidity .............................................................up to 95% non-condensing
Storage/Transport Temperature Range ..........................................-40° to +158°F (-40° to +70°C)
Storage/Transport Humidity .......................................................up to 95% non-condensing
Supplied Nebulizer Capacity ...................................................................10 ml (cc)
Supplied Nebulizer Nebulization Rate ..................................................0.15 ml/min or greater*
Supplied Nebulizer MMAD ..............................................................5 microns or less**
* Conditions may vary based on altitude above sea level, barometric pressure, and temperature.
**Value determined with minimum ow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there are no known risks associated with
disposal of the equipment.