Dickson KT6P Manual V3
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of ValidationReplaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
Dickson ensures that the Replaceable Sensors / Instruments listed below were
developed, tested and validated together.
Replaceable Sensor Model: R400
Instrument Model: KT6P1, KT6P2, KT6P5
Operation
OperationOperation
Operation: The Instruments and Replaceable Sensor models listed above were
specifically designed to work together. No other sensors will work with the Instrument
models listed. No other Instrument will work with the Replaceable Sensors listed. The
Replaceable Sensor must be connected to the Instrument for the Instrument to operate
properly.
Calibration
CalibrationCalibration
Calibration: Only the Replaceable Sensor is calibrated. The sensor, and all calibration
defaults and adjustments are stored on the Replaceable Sensor. Accuracy is strictly
controlled by the sensor – no adjustments are made by the Instrument. Readings are
sent from the Replaceable Sensor to the Instrument for storage purposes only.
Certificates of Calibration / NISTs are only supplied for the Replaceable Sensor for this
reason.
Recalibration
RecalibrationRecalibration
Recalibration: When a Replaceable Sensor is due for recalibration the existing
Replaceable Sensor can be replaced with a new Replaceable Sensor thus eliminating the
need to return the entire instrument for recalibration. Replaceable Sensors can be
returned for recalibration as well.