Infinity Delta/Delta XL/Kappa - Quick Reference Guide
Intended for quick reference only.
Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 19
© 2010 Draeger Medical, Inc.
Arrhythmia Problems
Monitor did not alarm
for arrhythmia
Cannot Learn Lead
Baseline Artifact
Possible Cause
Alarm not on, criteria
incorrect or HR source
incorrect. ALL Alarms
off chosen for default
setting.
Dominant normal
complex could not be
determined for Lead.
After 100 beasts,
monitor cannot deter-
mine the dominant
normal complex on any
lead selected for QRS
processing. Learning
continues.
Artifact is blocking
arrhythmia class-
ification. Patient’s
movement, shivering,
tremors. Bad electrode
contact. Excessive sig-
nal noise. Interference
from auxiliary equip-
ment.
Action
Access alarm limits,
turn on alarm with de-
sired limits check HR
source. Ensure that if
monitor is in OR mode
that the monitor is not
in ALL Alarms off (this
will have a timed
setting) mode or
BYPASS mode.
Check Leads. Choose
another lead for QRS
processing. Check
electrode preparation.
Check electrode
preparation. Calm
patient. Check
electrodes and reapply
if necessary. Ensure
that the patient’s skin
is properly prepped.
Isolate the patient form
auxiliary equipment.