What to do if Dräger Infinity Gamma IBP cannot zero?
- MmcookAug 8, 2025
If the Dräger Medical Equipment IBP cannot zero, consider these potential causes and solutions: * The transducer offset might be outside the zero balance range of +/- 190 mmHg. Verify the pressure cable connection to the monitor. * The signal may be too noisy, so keep all tubing motionless. * If the waveform is non-static (more than 3 mmHg variation in 3 seconds), check the stopcock and verify that the system is open to air completely. * If the monitor was unable to zero the transducer within 10 seconds, repeat the zeroing procedure. * If the issue persists, replace the transducer, as it may be faulty.