Page 9-6 Infinity Gamma Series VF4
Rate and Count
STEPS: Selecting Rate and Count Limits
1. Call up the Arrhythmia Setup table (see above).
2. Click on the limit in the
Rate or Count column for the
desired arrhythmia parameter.
3. Dial in the desired limit and click the knob.
NOTE: You can change limits only for VT, BRDY (rate only),
SVT, TACH, and PAUS (rate only). The displayed limits for RUN
and AIVR are informational.
Arrhythmia Alarms
STEPS: Turning Arrhythmia Alarms On/Off
1. Call up the Arrhythmia Setup table (see above).
2. Click on the alarm setting in the
Alarm column of the desired
arrhythmia parameter.
3. Dial in the desired setting (
ON/OFF * ) and click the knob.
NOTE: You cannot turn the alarms for asystole (ASY) and ven-
tricular fibrillation (VF) off.
WARNIN G: Certain ventricular tachycardia (VT) have
sinusoidal waveforms closely resembling those of ven-
tricular fibrillation (VF). Because of the similarity of these
waveforms, the monitor may classify such types of ven-
tricular tachycardia as ventricular fibrillation, the more
serious of the two conditions.