Default Settings and Biomedical Support
VF4 Infinity Gamma Series Page C-13
Startup Tests
The monitor runs internal tests continuously and upon power up
to check various functions, as well as the integrity of both ROM
and RAM memories. If any of the tests fail, the monitor resets
and displays a message (see the following table). The test results
are recorded in the Diagnostic Log.
If the same error is detected three times in succession within 10
minutes without any other intervening errors, the monitor stops
resetting and generates a loud, continuous error tone. If this hap-
pens, take the unit out of operation and call DrägerService.
NOTE: The monitor stores patient data and settings in an internal
battery backed-up memory. If any of the following messages appear
frequently, the internal battery needs to be changed.
Message Test
Default Patient Settings Restored Integrity of patient data settings.
Software Reset, See Diagnostic
Log
The software has previously detected and
logged a fault condition. See the Diagnostic log
for the exact cause of the reset.
Replace Battery Pack Battery <25%.
Check Internal Battery Assure integrity of patient data and/or settings.
Calendar Clock Reset Invalid clock setting at startup.
Low Battery Reset Voltage < 10V.
High Temperature Reset Internal temperature > 64°C.
NBP Fault Integrity of the NBP pneumatic constants.
Patient Data Erased Integrity of patient data.