Respiration Monitoring
VF4 Infinity Gamma Series Page 11-13
Relearning a Patient’s Respiration Pattern
When respiration monitoring is enabled and set to Auto, the mon-
itor takes about one minute to automatically learn the patient's
respiration pattern. You can initiate a new learning phase at any
time. When respiration monitoring is enabled and set to
Manual,
you must initiate the learning of the respiration pattern manually.
Whenever you change or reposition electrodes or if you notice
significant changes in the patient's breathing pattern, repeat the
learning process.
STEPS: Relearning a Patient’s Respiration Pattern
1. Click on the Rsp parameter box.
2. Click on
Relearn.
During the learning phase, the monitor displays the message Rsp
Relearning at the bottom of the screen and the
LRN label in the
Rsp parameter box. During the learning phase, the patient should
rest as quietly as possible and breathe normally.