Cleaning, Disinfecting, Sterilizing
VF4 Infinity Gamma Series Page B-3
Patient Cables
! Clean the patient cables with a gauze pad moistened with a
soap solution.
! Dry thoroughly with a lint-free cloth.
! To disinfect patient cables, wipe the cables with a gauze
moistened with diluted alcohol or a gluteraldehyde-based dis-
infectant.
! Dry thoroughly with a lint-free cloth.
! Sterilize the ECG patient cables at temperature below 70°C.
Reusable ECG Electrodes
! Clean grabber-wire clips regularly with a toothbrush.
! Remove any gel residue from the electrode by brushing it off
under running water.
! Clean the electrodes with a gauze moistened with a soap
solution.
! Dry thoroughly with a lint-free cloth.
! Disinfect the electrodes with a gauze moistened with diluted
alcohol or a gluteraldehyde-based disinfectant.
! Dry thoroughly with a lint-free cloth.